I  U  P  A  C






News & Notices

Organizations & People

Standing Committees

Divisions

Projects

Reports
..By Year
..By Division
..Other Committees
..Provisionals

Publications

Symposia

AMP

Links of Interest

Search the Site

Home Page

 

 

 

Pure Appl. Chem., Vol. 70, No. 10, pp. 2039-2045, 1998

PHYSICAL CHEMISTRY DIVISION

COMMISSION ON MOLECULAR STRUCTURE AND SPECTROSCOPY*

Specification of Components, Methods and Parameters in Fourier Transform Spectroscopy by Michelson and Related Interferometers

(Technical Report)

 

LIST A: For Fourier Transform Spectroscopy at Modest Resolution

> Default assumptions
> Group AI Components, methods and parameters that are rarely changed
> Group AII Components, methods and parameters that should be given in all papers
> Group AIII Components, methods and parameters that should be given when quantitative significance is claimed for absolute or relative intensities or lineshapes
> Notes about List A

Back to Index - Introduction - List B - List C


Default assumptions

The following components, methods and parameters are defined as the defaults. They are the most commonly used and need not be specified. Deviations from them should be specified.

  • A Michelson or related interferometer with rapid continuous mirror scan was used.
  • No optical filter was used. If an optical filter was used, specify its temperature and wavenumber range.
  • The apodized interferogram was zero-filled before Fourier transformation, so that the Fourier transform gave at least two spectral points in each resolution interval.
  • The interferogram was sampled at 0.6328 µm intervals (HeNe laser wavelength), i.e., the spectral bandwidth was in the range 0 to over 3 950 cm-1 but less than 7 899 cm-1.[see notes]
  • The source was a continuous incandescent source, e.g., a hot globar or wire, etc..

Group AI Components, methods and parameters that are rarely changed

These components, methods and parameters should be specified in an accessible journal in the first publication from an instrument. This report should be cited in subsequent papers together with any differences in these specifications.

  • The detector. For a cooled detector, its temperature and stated wavenumber range.
  • The beamsplitter used and the angle of incidence upon it.
  • Whether the interferogram was one-sided or two-sided.
  • The optical retardation velocity (for continuous scan).[see notes]
  • Any other signal-filters used, e.g., electronic or digital filters.

Group AII Components, methods and parameters that should be given in all papers

  • The sample, its temperature, its purity and how this was determined, the method of sampling.
  • The instrument manufacturer and model.
  • How water and carbon dioxide were removed from the optical path.
  • How the wavenumbers were calibrated.[see notes]
  • The maximum optical path difference in the interferogram, Xmax, or the nominal resolution, 1/Xmax AND the apodization function.[see notes]
  • Other instrumental factors that may reduce the instrumental resolution achieved.[see notes]
  • For step scan, the integration time at each point.
  • For continuous scan the number of scans signal-averaged in each interferogram.

Group AIII Components, methods and parameters that should be given when quantitative significance is claimed for absolute or relative intensities or lineshapes

  • How the phase correction was done, e.g., multiplicative or convolution method.
  • The type of spectrum calculated from the Fourier transform.[see notes]
  • The PNL measure of photometric non-linearity.[see notes]
  • Details of any arithmetic processing of the spectra after Fourier transformation and phase correction, with the exception of the use of a stated reference spectrum to calculate the transmittance spectrum, T, or -lg T.

Notes About List A

Useful references for the concepts and terms used are:

P. R. Griffiths and J. A. de Haseth. Fourier Transform Infrared Spectrometry, Wiley-Interscience, New York, 1986.
J. E. Bertie Vibrational Spectra and Structure, Ed. J. R. Durig, Elsevier, Amsterdam, 1985, 14, 221.

Note that it is not helpful to describe instrument parameters as, e.g., UPF=3. The physical significance of the parameter is required.

If the sampling interval is Dx, wavenumbers from 0 cm-1 to = 1/(2Dx) are sampled correctly. If Dx = 0.6328 µm, = 7 899 cm-1 and a spectral bandwidth from 0 to 7 899 cm-1 is sampled correctly. The spectral intensity may not extend up to 7 899 cm-1; if the signal and associated noise do not extend beyond 3 950 cm-1, a larger sampling interval may be used.

Back to List A

The optical retardation velocity, ORV, is the rate at which the optical path difference changes. It is twice the velocity of the moving mirror for a Michelson interferometer, and is 4 or 8 times the mirror velocity for other types of interferometer. Some manufacturers give the ORV indirectly. Thus, Bio-Rad gives the optical retardation velocity as a frequency, e.g., 5 kHz which means that the path difference changes by one HeNe laser wavelength 5000 times per second. From this, ORV = 5 000 x 0.6328 x 10-4 cm s-1 = 0.32 cm s-1.

Back to List A

Wavenumber calibration is necessary to obtain accurate wavenumbers from FT spectrometers, although they give extremely precise (reproducible) wavenumbers. Calibration is conveniently done with water vapor or indene-camphor-cyclohexane or polystyrene film (A. R. H. Cole, Tables of Wavenumbers for the Calibration of Infrared Spectrometers, 2nd edn, IUPAC, Pergamon Oxford, 1977). Most instruments give wavenumbers correct to ~±0.1 cm-1 if the laser wavenumber is entered as 15 798.002 cm-1.

Back to List A

Resolution here means instrumental spectral resolution, which is not uniquely defined. The nominal resolution is conveniently defined as the reciprocal of the maximum optical path difference, Xmax. (Xmax equals the maximum mirror displacement multiplied by 2 for a Michelson or by some other factor for certain related interferometers) The actual instrumental resolution is determined by Xmax and the apodization function, so Xmax and the apodization function should both be reported. If these can not be determined, the resolution claimed by the manufacturer for the settings used may be substituted, although this is not consistently defined. Note that the spectral resolution actually observed may be determined by the intrinsic width of the spectral line if this is greater than the instrumental resolution.

Two factors that can reduce the instrumental resolution,, below that determined by Xmax and the apodization function are the focal length, F, of the collimating mirror and the diameter, d, of the Jacquinot stop (the limiting aperture of the instrument, frequently the circular aperture at the first focus of the source, but the detector element in some instruments). = d2/(8F2). Typically F 25 cm and d 12 mm, from which 1 cm-1 between 3 000 and 4 000 cm-1. These factors should be described if resolution of 1 cm-1 or better is claimed above 3 000 cm-1.

Back to List A

The type of spectrum calculated from the Fourier transform may be either the magnitude spectrum, I = (C2 + S2), or the intensity spectrum, which is also called the phase-corrected amplitude spectrum, I = C cos d + S sin d ; here all terms change with wavenumber and I, C, S, and d are the calculated intensity, the cosine transform, the sine transform and the phase angle at the wavenumber in question. Random noise about a zero baseline is always positive in the (uncommon) magnitude spectrum and has random sign in the intensity spectrum.

Back to List A

The PNL measure of photometric non-linearity is defined here as: In a single beam spectrum, the ratio of the average baseline position to low wavenumber of the detector cut-off to the maximum signal, times 100%.

Back to List A


Back to Index - Introduction - List B - List C


Page last modified 13 July 1999.
Copyright ©1999 International Union of Pure and Applied Chemistry.
Questions or comments about IUPAC, please contact, the Secretariat.
Questions regarding the website, please contact Web Help.